Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Source facts
Action
Firm began notifying customers on July 30, 2026, via FedEx 2-day air.
Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
Initiated
2026-07-30
Root cause
Process change control
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Recall number
Z-3198-2026
Recall status
Open, Classified
CDRH record ID
221972
FDA classified
2026-09-16
Recall event number
99558
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.