Z-3200-2024: Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images

Recall record Z-3200-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3200-2024
Company in FDA record
Philips North America Llc
FDA classification date
Sep 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Source facts

Action
Philips issued URGENT Medical Device Correction dated 8/23/24 to US consignees via Fed'x. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take: A. Customers can continue using identified systems in accordance with the intended use. B. As a reminder, when using MR systems, follow the warnings listed in the IFU provided with your system: Refer to the warnings listed in the IFU section titled Moving the patient into the magnet bore: " Before starting a scan which initiates tabletop movement, always check that nothing can get caught or hit during tabletop movement. " Check patient, patient extremities, clothing, equipment and positioning aids. Guide cables and intravenous lines " Due care must be taken to ensure that no part of the patient's body, hair, clothing cables or infusion lines can be trapped or injured by any part of the equipment. C. If you experience resistance from the tabletop movement during patient scan, Stop-use immediately and contact your local Philips service representative. D. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. E. Please display attached Advisory with your system(s); ensure the notice is in a place likely to be seen/viewed by operators. F. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt via email to: Philips.Recall@philips.com. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 5. The actions planned by Philips to correct the problem Philips will contact you to schedule time for a Field Service Engineer (FSE) to visit your site to inspect the IRF carrier assembly and provide a correction if necessary (reference FCO78100588). Philips plans to start implementing corrections in Q4 2024. If you need a
Initiated
2024-09-03
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Recall number
Z-3200-2024
Recall status
Open, Classified
CDRH record ID
209925
FDA classified
2024-09-20
Recall event number
95289

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗