Z-3201-2018: Contour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system Con

Recall record Z-3201-2018 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3201-2018
Company in FDA record
Medtronic Inc.
FDA record creation date
Sep 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

CONTOUR PLUS LINK 2.4 meter with an incorrect unit of measure was included into meter kits and distributed. This meter contains the incorrect unit of measure mg/dL instead of mmol/L.

Source facts

Action
The firm, Medtronic, sent an "URGENT MEDICAL DEVICE RECALL- CONTOUR"PLUS LINK 2.4 Meter (MMT-1151SK - Batch# MD04M051P) Incorrect Unit of Measure" letter to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: A. Confirm if the Model (REF) of your meter is MMT-1151SK and the batch (LOT) number is MD04M051P. B. If the REF and LOT match this notice, stop using the CONTOUR"PLUS LINK 2.4 meter immediately. C. Call your local Helpline and let the customer service team know you received a notice stating you may have a CONTOUR"PLUS LINK 2.4 meter with the incorrect unit of measure. D. Return your recalled CONTOUR"PLUS LINK 2.4 meter to Medtronic by following the instructions provided by the Helpline representative. Medtronic will provide a free replacement. If you have any questions, contact the Program Manager, Field Corrective Action at 818-576-4700.
Initiated
2018-05-16
Root cause
Nonconforming Material/Component
Terminated
2020-06-30
Distribution
International Distribution to Slovakia only.
Recall number
Z-3201-2018
Recall status
Terminated
CDRH record ID
166154
FDA record created
2018-09-19
Recall event number
80685

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗