Open original FDA record ↗Recall record
- FDA identifier
- Z-3203-2011
- Company in FDA record
- The Anspach Effort, Inc.
- FDA classification date
- Sep 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Anspach Effort, Palm Beach Gardens, FL, is recalling their Single Use, Sterile Bone Cutting Burrs, Catalog #, S-1504TD, Lot #s D363039140. D323038110, D293037349 and Catalog # TURQ-434-4, Lot #s D343038845, D323038131, due to the burr being 2.0 mm longer than stated on the label. As a result, a hole drilled using this burr would be 2 mm deeper than the screw.
Source facts
- Action
- Anspach Effort, Palm Beach Gardens, FL sent a Customer letter dated March 31, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to screen their inventory, contact Anspach Customer Support at (800) 327-6887 if any of the recalled product was found, and complete and return the attached reply form. If the product had been further distributed, they were asked that the recall information be forwarded as appropriate.
- Initiated
- 2011-03-17
- Root cause
- Mixed-up of materials/components
- Terminated
- 2011-10-11
- Distribution
- Worldwide Distribution -- USA (nationwide) including the states of CA, CO, IN, KY, MD and TX. and the countries of Germany, Japan, Taiwan, United Kingdom and Sweden.
- Recall number
- Z-3203-2011
- Recall status
- Terminated
- CDRH record ID
- 101910
- FDA classified
- 2011-09-14
- Recall event number
- 59182