Z-3206-2024: BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Mag

Recall record Z-3206-2024 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3206-2024
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Sep 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the sterile barrier to be compromised.

Source facts

Action
On 08/28/2024, the firm sent an "URGENT: Medical Device Product Removal" Letter via FedEx and/or email to customers informing them that BD has identified through internal inspections, that the packaging of the Magnum Biopsy Needle may have been damaged by indentations unintentionally made during manufacture, potentially compromising the sterile barrier that may be hard to detect by the end user. Customers are instructed to: 1.Discontinue use of product listed in the letter, and immediately dispose of all affected product remaining in their possession in accordance with their local facility s process. 2.Share this notification with all users within their facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, ensure that these entities are contacted and informed of the recall. 3.Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not customers have any of the product so that BD may acknowledge their receipt of this notification and subsequently process any required product replacement. 4.Indicate on the response form the quantity of product identified at their facility and confirm that this product inventory was disposed of. 5.If customers purchased this product from a distributor, contact their distributor for further instructions and credit resolution. Contact BD representative if assistance is needed with this process or contact the number below for further assistance: North American Regional Complaint Center for Recall questions, Product Complaints, Technical Questions - Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com. Recall Related Questions - BDRC13@BD.COM
Initiated
2024-08-02
Root cause
Packaging change control
Distribution
Worldwide - U.S. Nationwide distribution in the Puerto Rico and the countries of Canada, Pakistan, Peru, India, Malaysia, China, Brazil, Mexico, Japan, Chile, Australia.
Recall number
Z-3206-2024
Recall status
Open, Classified
CDRH record ID
209672
FDA classified
2024-09-20
Recall event number
95174

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗