- FDA identifier
- Z-3208-2024
- Company in FDA record
- Percussionaire Corporation
- FDA classification date
- Sep 27, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.
Source facts
- Action
- On September 3, 2024, Percussionaire Inc. issued a "Urgent Medical Device Field Safety Notice" to affected consignees. Percussionaire asked consignees to take the following actions: 1. Complete and Return Acknowledgement form (see Appendix 1) after reviewing and implementing the requested actions below by September 27, 2024. 2. Review updated instructions for proper in-line use (see Appendix 3) and updated IFU (see Appendix 4) 3. Print and post updated instructions for proper In-Line Valve use (Appendix 3) throughout your facility for adequate distribution of new information to users. 4. Communicate and train with newly presented instructions for proper use to primary users of IPV therapy, specifically for in-line use. 5. Review institutional protocols related to P5-TEE inline use and update as needed to conform with the updated instructions. 6. Re-label current product by placing the sticker provided by Percussionaire on each individual P5-TEE bag on stock (stickers to be provided via mail). 7. Report any adverse events to regulatory.percussionaire@sentec.com and/or to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 8. Report any quality problems experienced with the use of this product to Percussionaire/Sentec Customer Service department via email to FSCA@sentec.com.
- Initiated
- 2024-08-21
- Root cause
- Device Design
- Distribution
- US: Nationwide OUS: Israel Japan Russia Switzerland Canada Turkey
- Recall number
- Z-3208-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209800
- FDA classified
- 2024-09-27
- Recall event number
- 95239