Open original FDA record ↗Recall record
- FDA identifier
- Z-3209-2024
- Company in FDA record
- Microcare Medical
- FDA classification date
- Sep 23, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Contaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused product bloating/leaking, and if used during medical/dental procedures could result in urinary tract, respiratory, bloodstream infections; workers handling contaminated product/instruments also at risk
Source facts
- Action
- On 8/29/24, recall notices were mailed to customers who were asked to do the following: 1) This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. This recall should be carried out to the user level. 2) If your company repacked these medical devices into convenience kits or into other packaging, please contact your local FDA recall coordinator to evaluate the need for a new recall (https://www.fda.gov/safety/industry-guidance-recalls/ora-recallcoordinators). 3) Return partially-used or unused product to the recalling firm. 4) Complete and return the response form via email to qdc@microcare.com If you have any questions, contact the firm at 303-529-5937 or the email above.
- Initiated
- 2024-08-29
- Root cause
- Material/Component Contamination
- Distribution
- Worldwide - US Nationwide distribution in the states of MN, GA, PA, WA, CO, NY, MD, CA, MI, NE, UT, OK, WI, NJ, IL, ME, CT, KS, TX, NC, TN, SC, MO and the countries of Canada.
- Recall number
- Z-3209-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209670
- FDA classified
- 2024-09-23
- Recall event number
- 95172