Skip to content

Public FDA record

Z-3210-2024: AVS Anchor-C Cervical Cage System Instructions For Use (IFU)- eIFU (revision 5)-Indicated for anterior ce

Recall record Z-3210-2024 for Stryker Spine. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3210-2024
Company in FDA record
Stryker Spine
FDA classification date
Sep 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling Correction: Instructions For Use (IFU)-Updated electronic Rev 5, specific language for indications for use for the United States and Canada was inadvertently removed.

Source facts

Action
Stryker issued Medical Device Correction letter on 8/23/24 to Surgeons via Fed'X. Letter states reason for recall, health risk and action to take: Our records indicate that you are a new user of the AVS-Anchor C system since the eIFU was updated to Revision 5 in January 2023. Please review the Product Correction Notification and the updated eIFU (Revision 6). This communication is being made for notification purposes only; there are no product(s) to be returned for this matter. 2. Use the Business Reply Form to acknowledge receipt of this notification letter. 3. Return the enclosed Business Reply Form by email to: Spine-egulatoryActions@Stryker.com to confirm receipt of this notification. 4. If you desire additional training associated with these instructions, please contact our Regulatory Compliance team by calling 201.749.8090 or by email at Spine-RegulatoryActions@Stryker.com, or contact your local Sales representative. Completing this action by the target date September 13, 2024 and regret any inconvenience that may be caused
Initiated
2024-08-23
Root cause
Labeling Change Control
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-3210-2024
Recall status
Open, Classified
CDRH record ID
209874
FDA classified
2024-09-23
Recall event number
95274

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗