- FDA identifier
- Z-3212-2024
- Company in FDA record
- Centinel Spine, Inc.
- FDA classification date
- Sep 24, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
Source facts
- Action
- Customers were contacted via telephone beginning 7/18/24. The Centinel Quality team has identified a potential nonconformance on three specific lots in the prodisc c product family. We are requesting the return of the inventory that is assigned to you. We will provide replacement inventory proactively to support the business. Thank you for your support. In addition, an URGENT - MEDICAL DEVICE Recall Instructions for PRODISC C IMPLANTS notification letter was sent to the distributor. Required Action: Please retrieve the identified devices and return the affected devices to Centinel Spine Depth of Recall: This recall should be carried out to the hospital level, including Distributors and Consigned locations Please acknowledge that you have read the email and enable the Return Receipt. This is necessary to ensure traceability as part of the recall process. If you have any questions, please contact Jessica Staub 484.887.8803.
- Initiated
- 2024-07-18
- Root cause
- Process control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-3212-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209638
- FDA classified
- 2024-09-24
- Recall event number
- 95164