Z-3212-2026: CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24

Recall record Z-3212-2026 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3212-2026
Company in FDA record
ICU Medical, Inc.
FDA classification date
Sep 17, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.

Source facts

Action
On August 20, 2026, the firm began notifying affected customers via Urgent Medical Device Correction letters. Customers were informed that certain reservoirs may exhibit fluid leakage at the joint between the connector and the tubing. Customers were instructed to check all inventory locations for affected product and discontinue use. All affected products should be destroyed following their institution's destruction process. If destruction is not immediately possible at your facility, quarantine the products until disposal. Customers should share the notification with all potential users of the reservoirs, and notify any accounts to whom the affected products have been further distributed. The firm will provide replacement products to affected customers upon receipt of a completed Customer Response Form to certify product destruction.
Initiated
2026-08-20
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, LA, MI, MO, MS, TN, TX, VA, WI and the countries of Austria, Australia, Belgium, Costa Rica, Germany, Estonia, Spain, Finland, France, Great Britain, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, and Sweden.
Recall number
Z-3212-2026
Recall status
Open, Classified
CDRH record ID
222072
FDA classified
2026-09-17
Recall event number
99648

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗