Z-3214-2024: GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.

Recall record Z-3214-2024 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3214-2024
Company in FDA record
GE Healthcare
FDA classification date
Sep 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.

Source facts

Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/05/2024 by letter delivered via traceable means. The notice explained the issue, potential risk to the patient, and requested the following: "You can continue to use your device if it has not reached the maximum number of Ckeys. In order to determine the remaining number of available Ckeys: " If GE HealthCare is remotely monitoring your Centricity PACS system as part of an existing service agreement, the number of available Ckeys is assessed routinely. If your device is close to maximum capacity, the GE HealthCare support team will contact you to discuss next steps. " If GE HealthCare is not remotely monitoring your system, please contact the GE HealthCare support team who will help you check your Ckey status. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions above." "GE HealthCare will provide an updated manual with CKey monitoring instructions at no cost to you. A GE HealthCare representative will contact you with further information." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2024-08-05
Root cause
Device Design
Distribution
Worldwide distribution.
Recall number
Z-3214-2024
Recall status
Open, Classified
CDRH record ID
209804
FDA classified
2024-09-24
Recall event number
95243

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗