Skip to content

Public FDA record

Z-3214-2026: Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/03

Recall record Z-3214-2026 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3214-2026
Company in FDA record
Boston Scientific Corporation
FDA classification date
Sep 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.

Source facts

Action
Firm began notifying customers on August 24, 2026. Customers were instructed to immediately stop further use or distribution and segregate affected product. Customers should continue to follow the standard of care when monitoring patients treated with the product. Customers were advised to utilize both inner and outer packaging UPN codes when searching for affected product. Customers were provided with instructions to return product.
Initiated
2026-08-24
Root cause
Process control
Distribution
Worldwide - US Nationwide and territories of Puerto Rico and the countries of Albania, Algeria, Argentina, Australia, Austria, Barbados, Belgium, Brazil, Brunei Dar-es-S, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Poland, Portugal, Puerto Rico, Qatar, Rep N Macedonia, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Trinidad, Tobago, Tunisia, T¿rkiye, UAE, Ukraine, United Kingdom, USA, W. Bank Gaza Strip, Yemen.
Recall number
Z-3214-2026
Recall status
Open, Classified
CDRH record ID
222200
FDA classified
2026-09-23
Recall event number
99687

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗