- FDA identifier
- Z-3215-2026
- Company in FDA record
- Heidelberg Engineering GmbH
- FDA classification date
- Sep 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.
Source facts
- Action
- Firm began notifying US customers on August 6, 2026 via email. Instruction to Customers: There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose when the Flex Module is being pulled or pushed during movement or transportation, affecting the stability of the stand. As a result, the Flex stand may tip over if not correctly handled or supported by the operator. Users shall schedule a service appointment with your Heidelberg Engineering service partner as soon as possible so that the castor fastenings can be inspected and, if necessary, corrected.
- Initiated
- 2026-08-06
- Root cause
- Component design/selection
- Distribution
- Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam.
- Recall number
- Z-3215-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221928
- FDA classified
- 2026-09-18
- Recall event number
- 99595