Open original FDA record ↗Recall record
- FDA identifier
- Z-3221-2011
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Sep 16, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is the potential that the current may not be controlled correctly in examinations using the ECG modulation function, which adjusts the x-ray output in synchronization with the ECG waveform.
Source facts
- Action
- Toshiba America Medical Systems (TAMS) issued an Urgent: Medical Device Correction notification to consignees via USPS return receipt mail. The letter identified the affected product, the problem, and corrective action. The firm plans to install revised firmware to prevent the occurrence of the problem. A Toshiba Service Representative will contact customers to schedule an appointment for installation. The firm asked customers to stop using the Power Reduction option during ECG modulated scanning until the new firmware has been installed. Customers are to share the information in the letter with all users, reviewing radiologist, and the clinical engineering or Biomedical group at their facility. Customers are also to complete and return the attached reply form via fax or e-mail to TAMS as proof of notification. Questions pertaining to the letter should be directed to 800-421-1968 or customers' local Toshiba Representative at 800-521-1968.
- Initiated
- 2010-12-08
- Root cause
- Software design
- Terminated
- 2013-03-18
- Distribution
- Nationwide Distribution.
- Recall number
- Z-3221-2011
- Recall status
- Terminated
- CDRH record ID
- 103009
- FDA classified
- 2011-09-16
- Recall event number
- 59519