Skip to content

Public FDA record

Z-3224-2026: PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD56

Recall record Z-3224-2026 for Maquet SAS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3224-2026
Company in FDA record
Maquet SAS
FDA classification date
Sep 18, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential for the device to detach at the junction between the spring arm and suspension arm and fall causing potential injury.

Source facts

Action
On August 7, 2026 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Identify any affected light(s) in your facility. Do not use affected light(s) until Getinge has completed the inspection and any required repairs. You may continue to use the room where the light is located. The light enclosure must remain closed; use an alternative light source as needed. 2. Forward this notification to anyone in your organization, or any organization Where the affected devices were transferred, who needs to be aware. 3. Complete and return the enclosed response form to recallresponses.qrc@getinge.com.
Initiated
2026-08-07
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-3224-2026
Recall status
Open, Classified
CDRH record ID
222211
FDA classified
2026-09-18
Recall event number
99696

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗