Open original FDA record ↗Recall record
- FDA identifier
- Z-3224-2026
- Company in FDA record
- Maquet SAS
- FDA classification date
- Sep 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The potential for the device to detach at the junction between the spring arm and suspension arm and fall causing potential injury.
Source facts
- Action
- On August 7, 2026 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Identify any affected light(s) in your facility. Do not use affected light(s) until Getinge has completed the inspection and any required repairs. You may continue to use the room where the light is located. The light enclosure must remain closed; use an alternative light source as needed. 2. Forward this notification to anyone in your organization, or any organization Where the affected devices were transferred, who needs to be aware. 3. Complete and return the enclosed response form to recallresponses.qrc@getinge.com.
- Initiated
- 2026-08-07
- Root cause
- Process control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-3224-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222211
- FDA classified
- 2026-09-18
- Recall event number
- 99696