Open original FDA record ↗Recall record
- FDA identifier
- Z-3228-2026
- Company in FDA record
- Medline Industries, LP
- FDA classification date
- Oct 1, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Source facts
- Action
- Firm initially notified customers November 26, 2025. On August 14, 2026, the firm sent a follow-up letter which provides additional product SKU - Item numbers for the electrical system safety risks and additional important information related to the previous notices. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 orrecalls@medline.com.
- Initiated
- 2026-08-14
- Root cause
- Device Design
- Distribution
- Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
- Recall number
- Z-3228-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222566
- FDA classified
- 2026-10-01
- Recall event number
- 99827