Z-3237-2017: AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ q

Recall record Z-3237-2017 for Radiometer America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3237-2017
Company in FDA record
Radiometer America Inc
FDA record creation date
Sep 29, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)

Source facts

Action
Radiometer sent a Safety Notice dated September 7, 2017, to all affected customers. The firm initiated their recall on 09/07/2017 by email. The letter requested the following: 1. Please check your inventory and analyzers for any boxes and ampoules of above mentioned QC. 2. Discard all Boxes and ampoules of this lot. 3. Complete the response form and return it to the number on the form. Please note: If you are not the end-user of the affected product, please ensure that this letter is distributed to the final end-user. Customers with questions were advised to contact their Radiometer representative.
Initiated
2017-09-07
Root cause
Under Investigation by firm
Terminated
2019-08-01
Distribution
Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and Finland.
Recall number
Z-3237-2017
Recall status
Terminated
CDRH record ID
158651
FDA record created
2017-09-29
Recall event number
78097

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗