Open original FDA record ↗Recall record
- FDA identifier
- Z-3238-2017
- Company in FDA record
- Bomimed
- FDA record creation date
- Sep 29, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for intermittent lighting (flickering) during use.
Source facts
- Action
- A letter was sent to all the consignees via mail dated July 13, 2017 informing of the recall due to Length variation of the battery contact inside the handle due to manufacturing/assembly variation may potentially cause an intermittent electrical connection resulting in flickering light. Complete the bottom section and return by email to regulatory@bomimed.com to confirm this notice has been received and understood. Check functionality of the light for each handle by engaging a blade and applying pressure to simulate use. Complete this task prior to use. Contact BOMImed for replacements if intermittent lighting is present. Replacement product orders will be organized in a manner to assist in covering all affected customers needs. Contac: Trina Friesen Quality Assurance & Regulatory Manager Tel: 800-667-6276 ext.234 Fax: 877-435-6984 regulatory@bomimed.com
- Initiated
- 2017-07-13
- Root cause
- Device Design
- Terminated
- 2021-04-01
- Distribution
- IL. PA. and WA CANADA
- Recall number
- Z-3238-2017
- Recall status
- Terminated
- CDRH record ID
- 158329
- FDA record created
- 2017-09-29
- Recall event number
- 77958