Open original FDA record ↗Recall record
- FDA identifier
- Z-3239-2017
- Company in FDA record
- Firefly Medical, Inc.
- FDA classification date
- Sep 7, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit.
Source facts
- Action
- The recall notification is expected to begin on or before 9/8/17 after pin cap components are received for the corrections. Instructions in the recall notification letter do not include return of recall product. They are to be corrected in the field. If any units are returned, they will be labeled and quarantined until corrected by Firefly Medical personnel or the sub-contracted manufacturer supplier. The recall notification letter will be sent certified letter with return receipt.
- Initiated
- 2017-08-24
- Root cause
- Device Design
- Terminated
- 2019-11-01
- Distribution
- United States, Australia, Belgium, Canada, China, Germany, Hong Kong, Indonesia, Israel, Netherlands, Singapore, UAE
- Recall number
- Z-3239-2017
- Recall status
- Terminated
- CDRH record ID
- 158512
- FDA classified
- 2017-09-07
- Recall event number
- 78019