Z-3240-2018: Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

Recall record Z-3240-2018 for Stryker GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3240-2018
Company in FDA record
Stryker GmbH
FDA record creation date
Sep 22, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.

Source facts

Action
Stryker contracted Third Party: Stericycle Contact Center, 6026 Lakeside Blvd., Indianapolis, IN 46278 notified affected hospitals via 2-day UPS (with return receipt) on July 26, 2018 All affected surgeon users notified via email with Request read Receipt and Request delivered Receipt options selected. Provided the following: if you have selected the Correct Axial First button, with multiple corrections per day, redo the correction plan by either of the following methods: a. Calculate the correction plan with one correction per day selected on the Correction Plan page or b. If more than one correction per day is desired, the Correction time may be overridden by entering the desired number of days in the Override correction time field on the Limiting Anatomical Structure (LAS) page. If the correction time is overridden, the correction plan will be accurate regardless of the number of corrections. Continue to evaluate the patients through routine follow-up radiographs for alignment and distraction.
Initiated
2018-07-26
Root cause
Under Investigation by firm
Terminated
2020-05-26
Distribution
Nationwide
Recall number
Z-3240-2018
Recall status
Terminated
CDRH record ID
166756
FDA record created
2018-09-22
Recall event number
80863

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗