Skip to content

Public FDA record

Z-3241-2018: Durex Real Feel Polyisoprene Condoms, 10 count and 24 count, SKU Numbers 02340-89456-14 and 02340-89457-0

Recall record Z-3241-2018 for Reckitt Benckiser LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3241-2018
Company in FDA record
Reckitt Benckiser LLC
FDA record creation date
Sep 24, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The products failed the filed specification of 1.0kPa for Burst Pressure.

Source facts

Action
On 7/23/18, an Urgent Medical Device Recall notification letter was distributed to customers. The letter instructed customers to do the following: 1. Examine your inventory in your distribution center immediately to determine if you have the specified batch in your warehouse inventory. 2. Segregate and quarantine the specified batch. 3. If you have product of the specified batch, contact our Stericycle representative immediately for a RETURN AUTHORIZATION number. a. Contact Information: Phone: 877-448-5308 Fax: 877-884-9410 4. Contact Stericycle Expert Solutions at 877-448-5308 for assistance with coordination of pickup and removal of product from your facility. If arranging pick up of product, freight will be prepaid by Stericycle. A pre-paid return label has been included for your convenience. Alternatively, you may arrange shipment of product directly to Stericycle at the following address: Attn: Event 4241 2670 Executive Drive, Suite A Indianapolis, IN 46241 Dock Hours are SAM -4:30PM Monday- Friday Please call to schedule a dock appointment 5. Finally, notify Stericycle representative via email RB4241@stericycle.com that you have: a. Checked your inventory in your distribution centers and confirm you have NO product of the specified batch b. You have collected all units in your inventory and product has been shipped back in accordance with the instructions from your Stericycle representative. On 7/26/18, a second Urgent Medical Device Recall notification letter was sent to customers which included two additional batches included in the recall. This letter included the same customer instructions as the 7/23/18 letter.
Initiated
2018-07-23
Root cause
Under Investigation by firm
Terminated
2020-03-05
Distribution
US Nationwide.
Recall number
Z-3241-2018
Recall status
Terminated
CDRH record ID
166702
FDA record created
2018-09-24
Recall event number
80841

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗