Screws on a Terumo Advanced Perfusion
System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system
Source facts
Action
Terumo CVS issued letter on 8/2/818 via Fed'X 2 day air to inform users of the problem for the Power Manager boards with unverifiable or out of specification screws, potential hazard and field correction. A Terumo Field Service Representative will contact users to schedule the field correction activities. Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET
Initiated
2018-08-28
Root cause
Process control
Terminated
2020-04-23
Distribution
Nationwide
Foreign:
Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam
Recall number
Z-3243-2018
Recall status
Terminated
CDRH record ID
167119
FDA record created
2018-09-25
Recall event number
80957