Skip to content

Public FDA record

Z-3244-2018: G8 Automated HPLC Analyzer - 723G8 Product Usage - In vitro diagnostic use for the quantitative measurem

Recall record Z-3244-2018 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3244-2018
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Sep 25, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Analyzers were distributed with software which lacks a 510K

Source facts

Action
On July 13, 2018, Tosoh Bioscience started to telephone customers to set up on-site visits. On July 16, 2018, Tosoh Bioscience began to issue URGENT MEDICAL DEVICE RECALL notices to customers via hand delivery. Immediate Actions to be taken by the Customer/User: - Thoroughly review the content of this letter with the Tosoh field service representative. - Allow the Tosoh field representative to verify the software version currently on your analyzer - If the software is v5.23, your analyzer needs no further action - If the software is v5.24, allow the Tosoh field representative to download all patient results from your analyzer, remove software version v5.24, and upload software version v5.23. - Continue to use your analyzer and report results with software v5.23 once these activities have been completed. - Complete and return the Acknowledgement Form to the Tosoh field representative. - File the Urgent Medical Device Recall notice with your laboratory records and forward the information to others in your laboratory. If you have obtained unexpected test results or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette OConnell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. She will be available to answer any questions Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Initiated
2018-07-13
Root cause
No Marketing Application
Terminated
2020-07-08
Distribution
US Nationwide Distribution in states of - CA, OK, VA 7 WI
Recall number
Z-3244-2018
Recall status
Terminated
CDRH record ID
165791
FDA record created
2018-09-25
Recall event number
80585

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗