Z-3245-2018: DX-D 600 Product Usage: The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging syste

Recall record Z-3245-2018 for Agfa N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3245-2018
Company in FDA record
Agfa N.V.
FDA classification date
Sep 25, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A supplier for the Agfa system noted a potential for the steel support cables of a similar device to break. This led to a mandatory review of all systems, changing the pulley and cables if any deterioration in the cables or irregular or excessive pulley wear. If the two cables that support the equipment to its roof anchor break, the equipment might fall and cause harm to the patient, or others.

Source facts

Action
On August 15, 2018, an "URGENT FIELD SAFETY NOTICE" letter was emailed to the US and Canadian consignees. The letter described the product issue: Breakage of the steel cables that support the equipment to its roof anchor may cause the equipment to fall down and potentially harm the patient, user, or third parties. Actions undertaken by Agfa NV: Soon Agfa NV or its representative will visit your site to do a verification/inspection of your unit related to this topic. Despite the fact that this issue potentially affects older systems with high workload, Agfa NV or its representative will verify/inspect all DX-D 600 systems. Recommended actions to be taken by the customer: If you notice an abnormal displacement in the vertical axis or any suspicious noise during the vertical movement, please notify your Agfa NV Service contact at once. Customers were asked to complete the attached feedback form to confirm receipt.
Initiated
2018-08-15
Root cause
Component design/selection
Terminated
2020-06-23
Distribution
Worldwide Distribution - US Nationwide in the states of CA, DC, FL, IL, IN, KS, LA, MA, MN, MO, NV, NY, OH, OR, SC and TN. Foreign distribution to Canada.
Recall number
Z-3245-2018
Recall status
Terminated
CDRH record ID
167342
FDA classified
2018-09-25
Recall event number
81046

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗