Z-3247-2024: BD Phoenix 100-Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST)

Recall record Z-3247-2024 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3247-2024
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Sep 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,

Source facts

Action
BD issued URGENT: Medical Device Product Correction letter to Distributors and End-users. Letter states reason for recall, health risk and action to take: 1. Immediately inspect your facility to identify if you have the affected product listed above. Reference Attachment 1 to confirm software version. 2. Ensure the contents of this notification are read and understood by those within your organization. 3. Continue use of the instrument. However, confirmatory testing should be performed on any gram-negative organism identified when there is laboratory evidence to suspect E. coli. until BD is able to perform the remediation to correct the issue. 4. It is not necessary to review previous test results and no additional clinical actions are recommended. 5. Share and post this customer letter with the users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness. 6. Complete the attached Customer Response Form and return to the BD contact, whether or not your you have the affected product noted on the form, so that BD may acknowledge your receipt of this notification per FDA requirements. BD has identified the root cause and will implement appropriate corrective actions to prevent recurrence of this issue. BD will contact your facility to schedule a service visit to provide an update to the PUD (Phoenix Update Data) and software to correct this issue. North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Complaints when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com Product Complaints, Technical Questions Recall related questions: BDRC24@bd.com
Initiated
2024-08-01
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Dominican Republic Ecuador, Egypt, El Salvador, Ethiopia, France, French Guiana, Georgia, Germany, Ghana, Greece, Guatemala, Guinea, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Mali, Mauritania, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Syrian Arab, Republic Taiwan, Tanzania, United Republic of Thailand, Timor-Leste, Trinidad And Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, Vietnam, Zambia, Zimbabwe.
Recall number
Z-3247-2024
Recall status
Open, Classified
CDRH record ID
209601
FDA classified
2024-09-24
Recall event number
95145

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗