- FDA identifier
- Z-3248-2018
- Company in FDA record
- Swissray Medical
- FDA classification date
- Sep 25, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Possible injury due to movement of the arm, calibration loss and communication loss.
Source facts
- Action
- November 2016, Swissray Medical issued official communication to customers as a Action Notification Report. The manufacturer, Swissray Medical AG, distributed the Work Instructions for the FCO and the Action Notification Report Form to all distributors and service entities who would be performing corrective actions in the field. The Swissray Field Engineer then presented the Report to the customer (end user facility) and communicated with the customer in-person while performing the corrective action on-site. When corrective actions were completed, both the customer and service personnel then sign the Report to confirm that the customer was informed about the corrective action and that corrective action was completed.
- Initiated
- 2016-08-09
- Root cause
- Device Design
- Distribution
- US Nationwide
- Recall number
- Z-3248-2018
- Recall status
- Open, Classified
- CDRH record ID
- 167341
- FDA classified
- 2018-09-25
- Recall event number
- 81045