Open original FDA record ↗Recall record
- FDA identifier
- Z-3249-2026
- Company in FDA record
- Philips Ultrasound, LLC
- FDA classification date
- Sep 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
Source facts
- Action
- On August 7, 2026, Philips issued a Urgent Medical Device Correction Notification to affected consignees. Philips requested consignees take the following actions: 1. Due to possibility of a weak PW Doppler Signal on EPIQ Elite running 14.0 SW, customers are advised not to use C5-1 Abd Vasc Tissue Specific Preset (TSP). Please use a different TSP (e.g., Abd Gen, Abd Renal) within C5-1. Please note that this issue (i.e., Weak PW Doppler signal) is restricted to the Abd Vasc TSP with C5-1 on EPIQ Elite running SW 14.0 only. 2. Otherwise, you may continue to use your device in accordance with the IFU. 3.Circulate this notice to all users of the device so that they are aware of the product issue and associated hazard / harm. 4. Post this notice near the affected unit(s) for ease of reference 5. Please return the completed response form at the end of this notice, indicating that information on this issue has been read and understood. Philips asks that you return the completed response form no later than 30 days from receipt of this letter.
- Initiated
- 2026-08-07
- Root cause
- Software Design Change
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan , United Kingdom, Vietnam.
- Recall number
- Z-3249-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222248
- FDA classified
- 2026-09-18
- Recall event number
- 99706