Open original FDA record ↗Recall record
- FDA identifier
- Z-3250-2026
- Company in FDA record
- Becton, Dickinson and Company
- FDA classification date
- Sep 21, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
Source facts
- Action
- On 08/11/2026, an "URGENT: Medical Device Correction" letter was email or FedEx'd informing customer that, based on customer feedback, BD has identified that during manufacturing, the laterality identification labels may be missing or inaccurately placed on the device. There is potential that the labels on the green cables may have been switched, with the channel 1 cable labeled as channel 2, and the channel 2 cable labeled as channel 1. This will lead to a mismatch as to the left and right values displayed on the monitor and what the labels indicate on the cables. Customers are instructed to check devices in clinical use and in inventory: 1. Verify for the presence of the flag, sensor, and channel labels on the device. 2. Verify that the flag and sensor labels correspond to the correct channel label on the housing (See picture included in the customer notification for correct label placement). 3. If either of the labels are missing, or the flag label and/or the sensor label do not match the channel label on the housing, contact Technical Support for further assistance. If the flag and sensor labels correctly correspond to the channel labels, as seen in the pictures, there is no further action required. 4. Provide a copy of this notification to any customers and/or facilities you may have distributed devices to. 5. Post this notice on or segregate the impacted unit(s) until correction is completed. 6. Complete the attached Customer Response Form and return it to the BD contact noted on the form whether or not you have any of the impacted products so that BD may acknowledge your receipt of this notification. For questions for assistance: Recall Customer Response Forms - Fax: 1-312-949-0410; Email: BDRC20@bd.com Technical/Product Complaints - Phone: 1.800.822.9837 Email: SMB-Complaints_US_APM@bd.com
- Initiated
- 2026-08-11
- Root cause
- Labeling mix-ups
- Distribution
- Worldwide - U.S Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Grand Cayman.
- Recall number
- Z-3250-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221435
- FDA classified
- 2026-09-21
- Recall event number
- 99370