Open original FDA record ↗Recall record
- FDA identifier
- Z-3251-2018
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA record creation date
- Sep 26, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue.
Source facts
- Action
- On 06/01/2018, Urgent Medical Device Correction letters and Consignee Response Forms were mailed to customers via overnight mail service. Customers were provided with additional information intended to supplement the system manual and instructed to call the firm at 1-888-826-5603 if they had any questions.
- Initiated
- 2018-06-01
- Root cause
- Under Investigation by firm
- Terminated
- 2022-02-02
- Distribution
- U.S.: MN, NY, PA, FL, AZ, MA, VA, CA, CO, AL, WI, TX, MO, DC, OH, NH, GA, SC, NJ, WA, CT, MI, OK, IN, MD, IL, KY, LA, OR, RI, UT, AR, MS, TN, NC Foreign: United Kingdom, Germany, Switzerland, Netherlands, Israel
- Recall number
- Z-3251-2018
- Recall status
- Terminated
- CDRH record ID
- 165134
- FDA record created
- 2018-09-26
- Recall event number
- 80308