Open original FDA record ↗Recall record
- FDA identifier
- Z-3253-2026
- Company in FDA record
- Criticare Technologies, Inc.
- FDA classification date
- Sep 22, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
Source facts
- Action
- Firm posted its initial notice on May 6, 2026. A follow-up letter was sent on August 31, 2026 with additional instructions. Customers were instructed to do the following: 1. Be aware of the Checkline Error alarm. If it appears on your CO2 section at startup, follow the steps provided in the letter. 2. Verify your backup plan. Confirm that your practice has access to an alternate CO2 monitoring source. 3. Retain the letter and note your unit s serial number for your records. 4. Contact the firm if you have any questions.
- Initiated
- 2026-05-06
- Root cause
- Process design
- Distribution
- US Nationwide distribution including in the states and territory of AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and Puerto Rico.
- Recall number
- Z-3253-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221806
- FDA classified
- 2026-09-22
- Recall event number
- 99545