Open original FDA record ↗Recall record
- FDA identifier
- Z-3258-2011
- Company in FDA record
- Angiodynamics, Inc.
- FDA classification date
- Sep 23, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some units of Nanoknife have the potential to deliver a shortened pulse width without indicating an error. The affected units contain a "GOMe" circuit board which may cause the generator to drift from the calibration points set at the time of manufacture.
Source facts
- Action
- AngioDynamics, Inc. sent a "NOTIFICATION OF MEDICAL DEVICE CORRECTION" letter dated April 4, 2011 to all affected customers. The letter describes the product, problem, and action to be taken. The letter instructs customers to review the measurement graphs (displayed on generator) after each delivery of energy, and to contact AngioDynamics Customer Service at 1-800-772-6446 if shortened pulses are detected. Contact your local sales representative or call Customer Service for any additional concerns or questions regarding this notice.
- Initiated
- 2011-04-04
- Root cause
- Component design/selection
- Terminated
- 2016-01-29
- Distribution
- Worldwide Distribution--USA (nationwide) and the countries of Australia, Germany, Israel, Italy, and Spain,
- Recall number
- Z-3258-2011
- Recall status
- Terminated
- CDRH record ID
- 103817
- FDA classified
- 2011-09-23
- Recall event number
- 59825