Z-3262-2024: PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is inten

Recall record Z-3262-2024 for Fisher & Paykel Healthcare, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3262-2024
Company in FDA record
Fisher & Paykel Healthcare, Ltd.
FDA classification date
Sep 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

Source facts

Action
On 08/13/2024, the firm sent an "Urgent: Medical Device Correction" Letter via email, phone, letter or face to face communication to inform customers that due to a software issue with the Airvo 3 (software version 1.5.1 or earlier) when the flow alignment alarm 3.2.2 occurs, Airvo 3 devices set up with high pressure oxygen the device will deliver room air only. Patients may experience oxygen desaturation that could lead to hypoxia. For Distributors - Actions required for affected products are: Step1-Identify affected products using Model Number reference (REF), Serial Number (SN) and Manufacturing Date (YYYY-MM-DD). If the device is still within your inventory, put the product in quarantine to prevent ongoing distribution until Fisher&Paykel (F&P) Healthcare has contacted you with next steps. Step2-Complete and return the attached response form Step3-Await instructions from Fisher&Paykel Healthcare For healthcare facilities: Until the software is updated, the Airvo 3 device can continue to be used by your customers. When using the device, all instructions, including warnings and cautions in the Airvo 3 User Manual must be followed, particularly those in Sections 1 and 2. If alarm 3.2.2 occurs, follow the onscreen instructions. For questions/assistance - contact F&P Field Action Team via email at FieldAction.USA@fphcare.com or directly at +1 (800) 446 3908 ext 5003 or +1 (949) 453 4000 ext 5003.
Initiated
2024-08-07
Root cause
Software change control
Distribution
U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided
Recall number
Z-3262-2024
Recall status
Open, Classified
CDRH record ID
209671
FDA classified
2024-09-25
Recall event number
95173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗