Open original FDA record ↗Recall record
- FDA identifier
- Z-3263-2011
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Sep 19, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because Beckman Coulter identified unexpected high particle counts in the lots of DxH Cell Lyse listed. These lots may generate high background counts; could result in a falsely elevated Total Nucleated Count (TNC) of up to 50 cells/L; or may also result in recovery of TNC values that were higher than expected for Level 1 of the Body Fluid Control.
Source facts
- Action
- The recall communication was initiated on 08/01/2011 with Beckman Coulter forwarding a Product Corrective Action (PCA) letter with attached PCA Response Form (via US Postal Service for US customers) to all customers, who purchased Coulter DxH Cell Lyse. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to contact Beckman Coulter for product replacement. In addition, customers were instructed to complete and return the enclosed PCA Response Form within 10 days. Customers with questions regarding this Product Corrective Action were instructed to call Beckman Coulter Customer Service at 800-526-7694 in the United States or Canada, or contact their Beckman Coulter representative.
- Initiated
- 2011-07-11
- Root cause
- Pending
- Terminated
- 2012-06-04
- Distribution
- Worldwide Distribution -- Australia, Belgium, Brunei Darussalam, Canada, France, Germany, Hong Kong, Hungary, India, Italy, Japan, Macao, Malaysia, Netherlands, Norway, Philippines, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and the United States.
- Recall number
- Z-3263-2011
- Recall status
- Terminated
- CDRH record ID
- 103461
- FDA classified
- 2011-09-19
- Recall event number
- 59726