Z-3265-2011: Powerheart Automated External Defibrillator (AED), G3 is indicated for emergency treatment of victims exh

Recall record Z-3265-2011 for Cardiac Science Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3265-2011
Company in FDA record
Cardiac Science Corporation
FDA classification date
Sep 19, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Cardiac Science Corporation (CSC) has identified two errors, 0x48 and 0x99, that have repeatedly occurred during a device self-test in the G3 installed base over the last 6 years. These self test errors cause the device to enter into a service required state. CSC remind owners/operators of the Powerheart AED devices to verify the "Rescue Readiness" of their equipment and follow maintenance pro

Source facts

Action
Cardiac Science began mailing out the recall notification letter on August 1, 2011, to all their US consignees via USPS first class mail. Cardiac Science's distribution partners will receive a notification letter and be provided with a template letter to send to their customers directly. CSC remind owners/operators of the Powerheart AED devices to check the status of of the AED every day to ensure it is Rescue Ready, and perform the regularly schedule maintenance as outlined in the Powerheart AED "Operator and Service Manual" provided with every AED. Customers can call Cardiac Science at 800-426-0337 or e-mail at aed205@cardiacscience.com.
Initiated
2011-06-30
Root cause
Device Design
Terminated
2014-02-13
Distribution
Worldwide Distribution - US (nationwide) This action applies only to domestic
Recall number
Z-3265-2011
Recall status
Terminated
CDRH record ID
102214
FDA classified
2011-09-19
Recall event number
59303

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗