Z-3270-2026: HeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US Device, US English -

Recall record Z-3270-2026 for HeartSine Technologies Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3270-2026
Company in FDA record
HeartSine Technologies Ltd
FDA classification date
Sep 25, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.

Source facts

Action
On 07/28/2026, Stryker sent several versions of an Urgent: Medical device recall letter that outlines two issues: 1) An issue with the membrane tail, an internal component, could cause an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs. 2) A limited number of HeartSine samaritan PAD devices were distributed without the required maintenance insert. This insert contains periodic maintenance instructions implemented as an additional risk control measure. While most devices contain the maintenance insert, this voluntary correction is being initiated to ensure all end users have access to the supplemental maintenance information. For the missing required maintenance insert (issue 2), Customers are instructed to: 1. Identify impacted devices by verifying if your device serial numbers are included in the list at the link https://heartsinerecall.com/HeartSineSamaritanPAD/LookUp. For this field action notice, select AED + Pad-Pak. 2. Download a copy of the missing maintenance insert when acknowledging the recall and store with your devices user manual until a replacement is received. For questions or concerns, contact Stryker Customer Service at +1 8007879537, option 2, from 8 AM to 7 PM Eastern Time, Monday to Friday, your Authorized Distributor, or HeartSine Technologies Technical Support at heartsinesupport@stryker.com.
Initiated
2026-07-27
Root cause
Process control
Distribution
U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Recall number
Z-3270-2026
Recall status
Open, Classified
CDRH record ID
218886
FDA classified
2026-09-25
Recall event number
98483

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗