Open original FDA record ↗Recall record
- FDA identifier
- Z-3270-2026
- Company in FDA record
- HeartSine Technologies Ltd
- FDA classification date
- Sep 25, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.
Source facts
- Action
- On 07/28/2026, Stryker sent several versions of an Urgent: Medical device recall letter that outlines two issues: 1) An issue with the membrane tail, an internal component, could cause an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs. 2) A limited number of HeartSine samaritan PAD devices were distributed without the required maintenance insert. This insert contains periodic maintenance instructions implemented as an additional risk control measure. While most devices contain the maintenance insert, this voluntary correction is being initiated to ensure all end users have access to the supplemental maintenance information. For the missing required maintenance insert (issue 2), Customers are instructed to: 1. Identify impacted devices by verifying if your device serial numbers are included in the list at the link https://heartsinerecall.com/HeartSineSamaritanPAD/LookUp. For this field action notice, select AED + Pad-Pak. 2. Download a copy of the missing maintenance insert when acknowledging the recall and store with your devices user manual until a replacement is received. For questions or concerns, contact Stryker Customer Service at +1 8007879537, option 2, from 8 AM to 7 PM Eastern Time, Monday to Friday, your Authorized Distributor, or HeartSine Technologies Technical Support at heartsinesupport@stryker.com.
- Initiated
- 2026-07-27
- Root cause
- Process control
- Distribution
- U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
- Recall number
- Z-3270-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218886
- FDA classified
- 2026-09-25
- Recall event number
- 98483