Skip to content

Public FDA record

Z-3280-2018: VITROS Sodium Slides, Catalog # 837 9034

Recall record Z-3280-2018 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3280-2018
Company in FDA record
Ortho-Clinical Diagnostics
FDA record creation date
Sep 28, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for positively biased results

Source facts

Action
On 05 May 2018, Customer letter (Ref. CL2018-123) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS customers, who have a analyzer capable of running VITROS Na+ slides, to inform them of the issue and to request that they immediately discontinue use of this VITROS Na+ Slides, GENs 8 and 13. Continued use of VITROS Na+ Slides, GENs14,16,17& 18 is acceptable provided the conditions detailed in the communication are met. Foreign affiliates were initially informed by email on 05 May 2018 of the issue and instructed to notify their consignees of the affected product with this issue and the requirement to immediately discontinue use of the VITROS Na+ slides GENs 8, 13, 14, 16, 17 and 18. Due to the amount of time required to ship replacement product to some worldwide locations, a revised communication to the foreign affiliates occurred on 08 June 2018. Due to the health risk associated with potentially leaving the laboratory without a Na+ method, the revised letter instructs customers that they may continue to use GENs 14, 16, 17 and 18 until replacement product is received.
Initiated
2018-05-05
Root cause
Nonconforming Material/Component
Terminated
2021-02-19
Distribution
US Nationwide, Bermuda, Austria, Brazil, Canada, Colombia, China, India, Japan, Mexico, Singapore, UK, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium and the Netherlands
Recall number
Z-3280-2018
Recall status
Terminated
CDRH record ID
165492
FDA record created
2018-09-28
Recall event number
80436

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗