Component defect. Philips Medical decided to conduct a recall field /correction (rework) of the mounting key locking threaded insert of the Detector Module system (DMS) bolt on one (1) Brilliance 64 CT X-ray unit. The firm initiated this action following Philip's receipt of the customer's complaint reporting that the retaining inserts were pulling out of the rotor during the removal of the UDMS
Source facts
Action
Philips sent a Field Change Order (FCO) 72800519 on June 13, 2011 to the affected customer. A Field Service Engineer visited the customer site and installed a hardware upgrade to the affected Brilliance 64 device system (Serial Number: 9592). The correction action was completed on August 9, 2011.
For questions regarding this recall call 1-440-483-7000.
Initiated
2011-06-02
Root cause
Other
Terminated
2013-03-15
Distribution
Nationwide Distribution including PA
Recall number
Z-3281-2011
Recall status
Terminated
CDRH record ID
102462
FDA classified
2011-09-27
Recall event number
59383