Skip to content

Public FDA record

Z-3297-2026: Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, GYN LAPAROSCOPY NEW HAN , C

Recall record Z-3297-2026 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3297-2026
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Sep 25, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.

Source facts

Action
An URGENT: PRODUCT CORRECTION notification letter dated 8/18/26 was sent to customers. Presource has completed its investigation regarding this recall and determined that you may have received Presource kits containing the affected Stryker component(s) and lot number(s). Please reference the attached report identifying your facility s kits, along with the corresponding lot numbers that are affected by this action. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: gmb-FieldCorrectiveAction@cardinalhealth.com Your response to this recall notice is required Please fax this form to 614-652-9648 or email to GMB-fieldcorrectiveaction@cardinalhealth.com to acknowledge your receipt of the recall. Please respond even if you do not have recalled product on hand.
Initiated
2026-08-18
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.
Recall number
Z-3297-2026
Recall status
Open, Classified
CDRH record ID
222440
FDA classified
2026-09-25
Recall event number
99768

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗