Open original FDA record ↗Recall record
- FDA identifier
- Z-3298-2026
- Company in FDA record
- Cardinal Health 200, LLC
- FDA classification date
- Sep 25, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
Source facts
- Action
- An URGENT: PRODUCT CORRECTION notification letter dated 8/18/26 was sent to customers. Presource has completed its investigation regarding this recall and determined that you may have received Presource kits containing the affected Stryker component(s) and lot number(s). Please reference the attached report identifying your facility s kits, along with the corresponding lot numbers that are affected by this action. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: gmb-FieldCorrectiveAction@cardinalhealth.com Your response to this recall notice is required Please fax this form to 614-652-9648 or email to GMB-fieldcorrectiveaction@cardinalhealth.com to acknowledge your receipt of the recall. Please respond even if you do not have recalled product on hand.
- Initiated
- 2026-08-18
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.
- Recall number
- Z-3298-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222441
- FDA classified
- 2026-09-25
- Recall event number
- 99768