Open original FDA record ↗Recall record
- FDA identifier
- Z-3301-2024
- Company in FDA record
- Cardinal Health 200, LLC
- FDA classification date
- Sep 26, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
Source facts
- Action
- The recalling firm issued letters dated 7/22/2024 via overnight mail on 7/24/2024. The letter informed the consignee the product labeling has been updated via electronic instructions for use for the affected product due to reported incidences of improper use of the device which can lead to breakage of the Salem Sump Anti-Reflux Valve (ARV) and an increased risk to patients. To mitigate the risk of improper use, the instructions have been revised. The risk to health and actions required were included in the letter. The actions included: (1) Review your inventory for the affected product codes; (2) Communicate the change of the use instructions with all personnel that utilize the devices; (3) Post a copy of Attachment A (the revised IFU) and the recall notification in the storeroom and clinical areas; (4) Notify any customers to whom you may have distributed/forwarded affected product or to whom you intend to distribute/forward product) regarding this medical device product correction and share a copy of the notice; (5) Return the enclosed acknowledgment form via FAX or email.
- Initiated
- 2024-07-24
- Root cause
- Labeling Change Control
- Distribution
- Distribution was nationwide, including Puerto Rico and Guam. There was also government/military distribution. Foreign distribution was made to Chile.
- Recall number
- Z-3301-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209748
- FDA classified
- 2024-09-26
- Recall event number
- 95214