Skip to content

Public FDA record

Z-3302-2024: (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in.

Recall record Z-3302-2024 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3302-2024
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Sep 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.

Source facts

Action
The recalling firm issued letters dated 7/22/2024 via overnight mail on 7/24/2024. The letter informed the consignee the product labeling has been updated via electronic instructions for use for the affected product due to reported incidences of improper use of the device which can lead to breakage of the Salem Sump Anti-Reflux Valve (ARV) and an increased risk to patients. To mitigate the risk of improper use, the instructions have been revised. The risk to health and actions required were included in the letter. The actions included: (1) Review your inventory for the affected product codes; (2) Communicate the change of the use instructions with all personnel that utilize the devices; (3) Post a copy of Attachment A (the revised IFU) and the recall notification in the storeroom and clinical areas; (4) Notify any customers to whom you may have distributed/forwarded affected product or to whom you intend to distribute/forward product) regarding this medical device product correction and share a copy of the notice; (5) Return the enclosed acknowledgment form via FAX or email.
Initiated
2024-07-24
Root cause
Labeling Change Control
Distribution
Distribution was nationwide, including Puerto Rico and Guam. There was also government/military distribution. Foreign distribution was made to Chile.
Recall number
Z-3302-2024
Recall status
Open, Classified
CDRH record ID
209872
FDA classified
2024-09-26
Recall event number
95214

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗