Z-3303-2024: Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip.

Recall record Z-3303-2024 for Healgen Scientific Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3303-2024
Company in FDA record
Healgen Scientific Llc
FDA classification date
Sep 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.

Source facts

Action
On 08/22/2024, the firm sent an initial "URGENT MEDICAL DEVICE (IVD) RECALL" Letter to customer to inform them that, Urinary Reagent Strip products have shown reduced sensitivity and potential false negative results for Protein test. An updated communication will be issued on 08/28/2024. Customers are instructed to: -Immediately examine their inventory and quarantine products subject to recall. -Stop the sale of this product immediately. -Stop or cease use of the products immediately. -Recall products in the market. -Ship recalled products to Healgen's warehouse in Houston within 1 month of this notice (OR dispose of the product and provide proof of destruction (records, pictures, etc.) to Healgen.) -Use Appendix II to record the inventory report of the affected lots and send it back to Healgen. -Inform the end-user or sub-distributor at once of this recall and the required actions. Customers may request the unaffected replacement products from their local Healgen sales at no charge. Contact QA of Healgen Scientific, LLC at quality@healgen.com or +1-713-733-8088 (Toll-Free) if you have any additional questions or concerns regarding this notification.
Initiated
2024-08-22
Root cause
Nonconforming Material/Component
Distribution
U.S. Nationwide distribution in the states of CA, GA and TX.
Recall number
Z-3303-2024
Recall status
Open, Classified
CDRH record ID
209812
FDA classified
2024-09-27
Recall event number
95248

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗