Z-3312-2024: DIVA 24 Inch Widescreen LCD Touch Display - Used with the Philips PIC iX system central monitoring system

Recall record Z-3312-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3312-2024
Company in FDA record
Philips North America Llc
FDA classification date
Sep 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a patient condition which prompted the alarm to generate. Although unlikely, this could potentially result in a patient harm.

Source facts

Action
Philips North America issued Important Product Information letter CR# 2023-CC-HPM-044, dated 2/15/24 to US consignees. Philips Market Organizations distribute the letters outside of the U.S. Letter states reason for recall, health risk and action to take: Actions that should be taken by the customer / user in order to prevent risks for patients or users 1) Please be aware of the alarms that remain active even though there is an indicator that alarm has been silenced. 2) Look for the water effect on the screen edges, if these exist, take the device out of service and arrange for alternative monitoring device to be used. This notice should be passed on to all those who need to be aware within your organization or to any organization where affected devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will replace the malfunctioning DIVA 24 Inch Widescreen LCD Touch Display. If you need any further information or support concerning this issue, please contact your local Philips representative.
Initiated
2024-02-15
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of CA, IL, ME, NC, NJ, NM, NY, OH, VA, WA and the countries of Australia, Austria, Canada, France, Germany, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, United Kingdom.
Recall number
Z-3312-2024
Recall status
Open, Classified
CDRH record ID
209971
FDA classified
2024-09-30
Recall event number
95316

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗