Z-3320-2026: In-Line Valve, Product Number: P5-TEE

Recall record Z-3320-2026 for Percussionaire Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3320-2026
Company in FDA record
Percussionaire Corporation
FDA classification date
Sep 28, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Source facts

Action
On 8/18/2026, correction notices were emailed to customers and distributors who were informed of the following: The following warning will be added to associated Instructions for Use: Warning: Do not use IPV therapy in conjunction with volatile anesthetic delivery systems. The high-frequency percussive flow generated during IPV therapy may alter the controlled delivery of volatile anesthetic agents within the breathing circuit, resulting in unpredictable anesthetic agent concentrations that may result in serious patient harm. Note: Consult the anesthesia healthcare provider before disconnecting from the system if considering delivering IPV directly to the patient's airway. Customers were asked to do the following: 1) Print and post the new warning for applicable IPV devices throughout your facility to ensure adequate distribution of new information to users. 2) Reinforce adherence to the IPV User Guide and IPV Quick Reference Guide instructions, including ensuring communication and training primary users of IPV therapy, specifically those using it in-line with a ventilator, on the instructions and new warning. Distributors were asked to do the following: - Translate (if applicable) and send the Urgent Medical Device Field Safety Notice to affected end customers. -If the Distributor receives the signed acknowledgement letters from affected customers, please forward them and all related communication to FSCA@sentec.com. If there are any additional questions related to the updated instructions or other questions, please contact FSCA@sentec.com
Initiated
2026-08-18
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
Recall number
Z-3320-2026
Recall status
Open, Classified
CDRH record ID
222002
FDA classified
2026-09-28
Recall event number
99617

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗