Open original FDA record ↗Recall record
- FDA identifier
- Z-3326-2026
- Company in FDA record
- Micro Therapeutics, Inc.
- FDA classification date
- Sep 28, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to incorrect label/product dimensions. Product is supposed to be 3.00 mm in diameter, but unit is packaging was a 2.50 mm in diameter.
Source facts
- Action
- On 06/25/2026, an "URGENT: MEDICAL DEVICE RECALL" letter was sent to customers to inform them that, Medtronic has confirmed a labeling error involving implant diameter in the affected lot. Devices from Lot # D112860 labeled as PED2-300-12 (3.00 mm diameter, 12 mm length) contain 2.50 mm diameter implants instead of the labeled 3.00 mm diameter. Customers are instructed to: 1. Do NOT use any impacted Pipeline Flex Embolization Device with Shield Technology devices listed. Remove and quarantine all unused impacted product units in their inventory. 2. Return the impacted product(s) to Medtronic. Their Medtronic representative can assist in facilitating the return of product and/or identifying suitable replacement product as necessary. 3. Complete the attached Customer Confirmation Form and email it to neuro.quality@medtronic.com. 4. Review their records to identify any patients who may have been treated with devices from the affected lot and assess whether any follow-up is appropriate. 5. Share this recall communication within their organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Maintain a copy of this letter for their records and the records of their patients with Pipeline Shield. For questions - contact Medtronic at 1 (800) 633-8766 or (763) 514-4000 (Worldwide) or Email: rs.nvquality@medtronic.com
- Initiated
- 2026-06-25
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Nationwide distribution in the states of GA and TX.
- Recall number
- Z-3326-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222202
- FDA classified
- 2026-09-28
- Recall event number
- 99689