Z-3326-2026: Pipeline Flex Embolization Device with Shield Technology REF PED2-300-12 UDI-DI code: 00763000448806 he

Recall record Z-3326-2026 for Micro Therapeutics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3326-2026
Company in FDA record
Micro Therapeutics, Inc.
FDA classification date
Sep 28, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to incorrect label/product dimensions. Product is supposed to be 3.00 mm in diameter, but unit is packaging was a 2.50 mm in diameter.

Source facts

Action
On 06/25/2026, an "URGENT: MEDICAL DEVICE RECALL" letter was sent to customers to inform them that, Medtronic has confirmed a labeling error involving implant diameter in the affected lot. Devices from Lot # D112860 labeled as PED2-300-12 (3.00 mm diameter, 12 mm length) contain 2.50 mm diameter implants instead of the labeled 3.00 mm diameter. Customers are instructed to: 1. Do NOT use any impacted Pipeline Flex Embolization Device with Shield Technology devices listed. Remove and quarantine all unused impacted product units in their inventory. 2. Return the impacted product(s) to Medtronic. Their Medtronic representative can assist in facilitating the return of product and/or identifying suitable replacement product as necessary. 3. Complete the attached Customer Confirmation Form and email it to neuro.quality@medtronic.com. 4. Review their records to identify any patients who may have been treated with devices from the affected lot and assess whether any follow-up is appropriate. 5. Share this recall communication within their organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Maintain a copy of this letter for their records and the records of their patients with Pipeline Shield. For questions - contact Medtronic at 1 (800) 633-8766 or (763) 514-4000 (Worldwide) or Email: rs.nvquality@medtronic.com
Initiated
2026-06-25
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of GA and TX.
Recall number
Z-3326-2026
Recall status
Open, Classified
CDRH record ID
222202
FDA classified
2026-09-28
Recall event number
99689

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗