Z-3328-2026: Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/C

Recall record Z-3328-2026 for SPECTRUM MEDICAL SRL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3328-2026
Company in FDA record
SPECTRUM MEDICAL SRL
FDA classification date
Sep 29, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to blood-to-glycol leak in the heat exchanger.

Source facts

Action
On 07/17/2026, an "URGENT MEDICAL DEVICE RECALL" letter was sent via email to customers informing them that, Spectrum Medical S.r.l. has received customer complaints reporting a leak between the blood and glycol compartments of certain adult heat exchangers. Actions to be taken by customers: During the procedure, the user should monitor the heat exchanger for leakage, which would be evident by a change in colour in the glycol compartment. Below are images of the heat exchanger if a blood-to-glycol leak was to occur. When operating as intended, the glycol compartment should remain clear. If the colour of the glycol compartment changes, then a leak is present between the blood and glycol compartments. If leakage is noted between the blood and glycol compartments of the heat exchanger, the heat exchanger should be immediately replaced in the circuit with a new device. Please return the defective heat exchanger to Spectrum Medical. The contaminated heater-cooler device should also be immediately removed from clinical use and returned to Spectrum Medical for decontamination and inspection. Do not use the heater-cooler with another patient. If leakage has occurred, please contact a Spectrum Medical representative to coordinate device return. If leakage into the water compartment occurs when using a third-party heater-cooler, the customer should immediately remove the heater-cooler from clinical use and contact the heater-cooler manufacturer. Spectrum Medical suggests that for customers that use the heat exchanger in combination with a third-party heater-cooler to contact the manufacturer of the heater-cooler to pre-emptively discuss the steps that should be taken if the device is exposed to blood contamination. For questions or further assistance - contact Regulatory Affairs Specialist at cameron.schmitt@spectrummedical.com
Initiated
2026-07-17
Root cause
Process design
Distribution
US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.
Recall number
Z-3328-2026
Recall status
Open, Classified
CDRH record ID
221606
FDA classified
2026-09-29
Recall event number
99462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗