Skip to content

Public FDA record

Z-3330-2026: Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog N

Recall record Z-3330-2026 for Tandem Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3330-2026
Company in FDA record
Tandem Diabetes Care, Inc.
FDA classification date
Sep 29, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.

Source facts

Action
On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that an unexpected behavior was identified in the t:slim iOS mobile application version 2.13 made available through TestFlight. The issue involves users unable to pair a compatible FreeStyle Libre 3 Plus sensor after updating to version 2.13. If this occurs, the application will incorrectly display an "Incompatible Sensor" message and prevent sensor use. Customer are advised to: 1. Discontinue use of iOS t:slim mobile application version 2.13 2. Uninstall version 2.13 from their mobile device. 3. Return to the current production version of the t:slim mobile application (version 2.12.1) available on the Apple App Store. 4. Scan the compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting their next sensor session. For Questions - contact Tandum Customer Support at 1-877-801-6901 or email Techsupport@tandumdiabetes.com (24/7/365).
Initiated
2026-07-16
Root cause
Software Design Change
Distribution
US Nationwide distribution.
Recall number
Z-3330-2026
Recall status
Open, Classified
CDRH record ID
221918
FDA classified
2026-09-29
Recall event number
99589

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.