Open original FDA record ↗Recall record
- FDA identifier
- Z-3330-2026
- Company in FDA record
- Tandem Diabetes Care, Inc.
- FDA classification date
- Sep 29, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
Source facts
- Action
- On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that an unexpected behavior was identified in the t:slim iOS mobile application version 2.13 made available through TestFlight. The issue involves users unable to pair a compatible FreeStyle Libre 3 Plus sensor after updating to version 2.13. If this occurs, the application will incorrectly display an "Incompatible Sensor" message and prevent sensor use. Customer are advised to: 1. Discontinue use of iOS t:slim mobile application version 2.13 2. Uninstall version 2.13 from their mobile device. 3. Return to the current production version of the t:slim mobile application (version 2.12.1) available on the Apple App Store. 4. Scan the compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting their next sensor session. For Questions - contact Tandum Customer Support at 1-877-801-6901 or email Techsupport@tandumdiabetes.com (24/7/365).
- Initiated
- 2026-07-16
- Root cause
- Software Design Change
- Distribution
- US Nationwide distribution.
- Recall number
- Z-3330-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221918
- FDA classified
- 2026-09-29
- Recall event number
- 99589