Z-3332-2026: Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code

Recall record Z-3332-2026 for Beckman Coulter Mishima K.K.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3332-2026
Company in FDA record
Beckman Coulter Mishima K.K.
FDA classification date
Oct 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

Source facts

Action
On 08/03/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was sent via postal mail and email informing customers that CHEMISTRY ANALYZER AU480 with Windows 10 IoT Enterprise 2021 LTSC PCsmay output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. Customers are instructed to: 1. Be sure to create a new index once or more per day at an appropriate time. Refer to 5.1 Start the System of the IFU to complete this action. 2. Beckman Coulter recommends posting this letter on or near the affected analyzers. Resolution - Beckman Coulter has identified the root cause and will resolve the issue by applying a Software solution to the impacted systems. BEC service teamwill contact customers to schedule the Software solution application, and a Field Service Engineer (FSE) will perform the on-site work. For questions - contact Customer Support Center at: http://www.beckmancoulter.com
Initiated
2026-08-03
Root cause
Software design
Distribution
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
Recall number
Z-3332-2026
Recall status
Open, Classified
CDRH record ID
221917
FDA classified
2026-10-01
Recall event number
99588

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗