Skip to content

Public FDA record

Z-3333-2026: HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II tele

Recall record Z-3333-2026 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3333-2026
Company in FDA record
Karl Storz Endoscopy
FDA classification date
Oct 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to incorrect inside packaging label for eyepiece.

Source facts

Action
On 08/12/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was sent via email informing customers that, KARL STORZ distributed 6 HOPKINS Telescope 30, 4mm REF: 10005BA between November 24, 2025, and December 15, 2025, that contains an optic (eyepiece) that is incorrectly etched with REF: 10005AA, which is a HOPKINS Telescope 0, 4 mm. The optic (eyepiece) inside the packaging is in fact a REF: 10005BA HOPKINS Telescope 30, 4mm, as the outer packaging indicates. Customers are instructed to: 1. Immediately check for impacted serial numbers in inventory and quarantine the incorrectly etched devices for return to Karl Storz. 2. Pass on this Urgent Medical Device Correction to all users of the products listed below (Table 1 and Table 2) and all other persons who need to be aware within your organization. 3. If customers have or may have distributed the products listed, identify and promptly notify those recipients with a copy of this notice and response form. 4. Return the completed Customer Reply Form by E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. For questions - email to KSEAProductLabeling@karlstorz.com.
Initiated
2026-08-12
Root cause
Employee error
Distribution
U.S. Nationwide distribution in the states of AZ, CA and MO.
Recall number
Z-3333-2026
Recall status
Open, Classified
CDRH record ID
221914
FDA classified
2026-10-01
Recall event number
99586

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.