Open original FDA record ↗Recall record
- FDA identifier
- Z-3334-2026
- Company in FDA record
- Wright Medical Technology, Inc.
- FDA classification date
- Oct 1, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
Source facts
- Action
- On 8/13/2026, recall notices were mailed and emailed to customers who were asked to do the following: 1) Review the product issue and risks as communicated in this Recall Notice and communicate the issue as appropriate in your organization. 2) Sign and return the enclosed business reply form by email to memphis.fieldaction@stryker.com 3) Upon receipt of the completed business reply form, firm will arrange for product return. 4) If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. Inform firm if any of the subject devices have been distributed to other organizations, including contact details, so that firm can inform the recipients appropriately. b. If you are a distributor, notify your affected customers. For questions or concerns, contact the firm at memphis.fieldaction@stryker.com
- Initiated
- 2026-08-18
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
- Recall number
- Z-3334-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222085
- FDA classified
- 2026-10-01
- Recall event number
- 99651