Skip to content

Public FDA record

Z-3334-2026: CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System

Recall record Z-3334-2026 for Wright Medical Technology, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3334-2026
Company in FDA record
Wright Medical Technology, Inc.
FDA classification date
Oct 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.

Source facts

Action
On 8/13/2026, recall notices were mailed and emailed to customers who were asked to do the following: 1) Review the product issue and risks as communicated in this Recall Notice and communicate the issue as appropriate in your organization. 2) Sign and return the enclosed business reply form by email to memphis.fieldaction@stryker.com 3) Upon receipt of the completed business reply form, firm will arrange for product return. 4) If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. Inform firm if any of the subject devices have been distributed to other organizations, including contact details, so that firm can inform the recipients appropriately. b. If you are a distributor, notify your affected customers. For questions or concerns, contact the firm at memphis.fieldaction@stryker.com
Initiated
2026-08-18
Root cause
Process control
Distribution
US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
Recall number
Z-3334-2026
Recall status
Open, Classified
CDRH record ID
222085
FDA classified
2026-10-01
Recall event number
99651

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗